Improving Planar Bending Stability and Kink Resistance

CASE STUDY | CATHETER SHEATH

Optimising Steerable Catheter Performance: Improving Planar Bending Stability and Kink Resistance

How systematic structural optimisation helped address shape instability while improving anti-kinking performance, pushability and torque transmission.

>45°

Original planar deviation

<8°

Optimised planar deviation

>90°

Bending angle with lumen ovality maintained

THE CHALLENGE

When steerable catheter performance depends on precise mechanical control.

During design validation of a customer’s steerable catheter, the deflection segment exhibited radial deformation during steering. The bending plane also deviated by more than 45° from the pull-wire exit orientation, resulting in inaccurate tip steering and reduced device deliverability.

01

RADIAL DEFORMATION

The deflection segment experienced radial deformation during steering, affecting structural stability and lumen geometry.

02

PLANAR BENDING DEVIATION

The bending plane deviated by more than 45° from the pull-wire exit orientation, reducing steering accuracy and predictability.

THE SALT MEDICAL APPROACH

Re-engineering the catheter for improved bending stability

Salt Medical applied a systematic steerable catheter design optimisation approach, evaluating the shaft material architecture, reinforcement, braid design, and pull-wire positioning. The objective was to improve planar bending stability, reduce radial deformation, and maintain lumen integrity during high-angle catheter deflection.

Material & reinforcement architecture

Redesigned the outer shaft material architecture and incorporated nitinol coil reinforcement within the deflection segment to improve resistance to radial deformation.

Gradient braid architecture 

 

Replaced the uniform braid density with a gradient PPI design to improve deformation resistance while supporting controlled bending behaviour.

 

 

Braiding & pull-wire process optimisation

Redesigned the braiding die and optimised the spring configuration to improve pull-wire positioning and planar bending stability.

 

 

 

THE RESULT

Reducing planar deviation for more accurate steering

Following the engineering optimisations, planar bending deviation was reduced from more than 45° to within 8°. The optimised catheter demonstrated improved steering accuracy, greater resistance to compression, and maintained lumen ovality during bending angles greater than 90°.

≤8°

OPTIMISED PLANAR DEVIATION

Steering accuracy

Improved tip steering and more predictable catheter deflection.

Compression resistance

Improved resistance to radial deformation during steering.

Lumen integrity

Lumen ovality maintained during bending angles greater than 90°.

Device deliverability

Smooth device passage through the working lumen.

HAVE A SIMILAR CHALLENGE?

Let’s solve your next catheter performance challenge.

Salt Medical works with medical-device innovators across design, engineering and manufacturing challenges. Discuss your application with our team.

RELATED INSIGHTS

Explore more medical device engineering challenges

Following a systematic review of the application requirements, the Salt Medical team optimised the sheath structure using coiled spring reinforcement principles.